Specifications
Environmental
op e r a t i n g St o r a g e
te m p e r a t u r e 41 to 95° F (5 to 35° C) -4 to 140° F (-20 to 60° C)
re l a t i v e hu mi D it y 15 to 95% (non-condensing) 15 to 95% (non-condensing)
at m o S p h e r i c pr e S S u r e 101 to 77 kPa (0 - 2286 m / 0 - 7500 ft) n/a
Physical
Dimensions: 10” L x 7.5” W x 5.5” H (25 x 19 x 14 cm)
Weight: Approximately 3.7 lbs (1.67 kg)
Water Capacity: 350 ml at recommended water level
Standards Compliance
This device is designed to conform to the following standards:
IEC 60601-1 General Requirements for Safety of Medical Electrical Equipment
EN ISO 17510-1 Sleep Apnea Breathing Therapy Devices
EN ISO 8185 General Requirements for Humidification Systems
Electrical
AC Power Consumption: 100 – 240 VAC, 50/60 Hz, 2.7 A max.
Type of Protection Against Electric Shock: Class II Equipment
Degree of Protection Against Electric Shock: Type BF Applied Part
Degree of Protection against Ingress of Water: Device: Drip Proof, IPX1
Mode of Operation: Continuous
Electromagnetic Compatibility: The device meets the requirements of EN 60601-1-2, 2nd edition.
Fuses: There are no user-replaceable fuses.
Humidity
Humidity
min
Output: 10 mg H
2
O/L
Measured @ max flow, 35° C, 15% RH.
Pressure
Pressure Increments: 4.0 to 20.0 cm H
2
O (in 0.5 cm H
2
O increments)
Pressure Stability: < 10 cm H
2
O (≤0.5 cm H
2
O peak-to-peak)
≥ 10.0 to 20 cm H
2
O (≤1.0 cm H
2
O peak-to-peak)
Measured in accordance with EN ISO 17510-1, 2002 @ 1/3, 2/3, and Pmax with BPM set to 10, 15, and 20 BPM.
Maximum Flow: 35 LPM
Measured in accordance with EN ISO 17510-1, 2002 @ 1/3, 2/3, and Pmax with BPM set to 10, 15, and 20 BPM.
Control Accuracy
Pa r a M e t e r ra n g e ac c U r a c y
CPAP 4 to 20 cm H
2
O ± 0.5 cm H
2
O
Measured at the patient end of the circuit with a Whisper Swivel II exhalation device and no patient flow.
Noise
Sound Pressure Level: < 30 dB(A)
Measured in accordance with EN ISO 17510-1 @ 10 cm H
2
O at the patient circuit.