Respironics 920M Manual de Serviço Página 51

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 50
43
English
Troubleshooting Guide (continued)
Symptom Possible Cause Possible Solution
Erratic display
and/or yellow
perfusion LED
during
concurrent use of
electrosurgical
equipment (ESU)
ESU may be
interfering with
oximeter performance
Examine the patient:
move oximeter,
cables, and sensor as
far away from ESU as
possible or refer to the
ESU operator's
manual
Perfusion is
blinking yellow
with each pulse
Perfusion signal at
sensor site is
marginal
Examine the patient:
reposition sensor or
select alternate sensor
site
Unable to obtain
green perfusion
Low patient pulse
strength
Reposition sensor on
patient
Sensor site poorly
perfused
Sensor not correctly
positioned
Sensor attached too
tightly or tape or other
items are restricting
perfusion at sensor
site
Reapply sensor, select
alternate sensor site,
or remove restrictive
material from sensor
site
Circulation reduced
due to excess
pressure between the
sensor and a hard
surface
Allow sensor and
finger to rest
comfortably on surface
Excessive ambient
light
Reduce ambient light
Excessive patient
motion
Reduce patient motion
Sensor applied to a
polished fingernail
Remove fingernail
polish
Interference from:
arterial
catheter
blood pressure
cuff
electrosurgical
procedure
infusion line
Reduce or eliminate
interference
Vista de página 50
1 2 ... 46 47 48 49 50 51 52

Comentários a estes Manuais

Sem comentários